Development of a ¹²⁷Xe calibration source for nEXO
Jan 12, 2022·,,,,,
,,,,
B.G. Lenardo
C.A. Hardy
R.H.M. Tsang
J.C. Nzobadila Ondze
A. Piepke
S. Triambak
Ako Jamil
G. Adhikari
S. Al Kharusi
E. Angelico
et al.
Abstract
We study a possible calibration technique for the nEXO experiment using a ¹²⁷Xe electron capture source. nEXO is a next-generation search for neutrinoless double beta decay (0νββ) that will use a 5-tonne, monolithic liquid xenon time projection chamber (TPC). The xenon, used both as source and detection medium, will be enriched to 90% in ¹³⁶Xe. To optimize the event reconstruction and energy resolution, calibrations are needed to map the position- and time-dependent detector response. The 36.3 day half-life of ¹²⁷Xe and its small Q-value compared to that of ¹³⁶Xe 0νββ would allow a small activity to be maintained continuously in the detector during normal operations without introducing additional backgrounds, thereby enabling in-situ calibration and monitoring of the detector response. In this work we describe a process for producing the source and preliminary experimental tests. We then use simulations to project the precision with which such a source could calibrate spatial corrections to the light and charge response of the nEXO TPC.
Type
Publication
Journal of Instrumentation, 17, P07028 (2022)